Temporary Pacemakers

Across
DDD. Paces and senses both the atrium and the ventricle.
How the pacemaker sees the intrinsic rhythm. Measured in millivolts (mV)
These are on the patient's right
The amount of energy in milliamps(mA)sent to the atrium
The amount of energy in milliamps (mA) sent ot the ventricle
Single chamber pacing that paces and senses the atrium. From DDD mode,turn ventricular output (mA) to "0" and the mode changes to AAI
Single chamber pacing that paces and senses the ventricle. From DDD mode,turn atrial output (mA) to "0" and the mode changes to VVI.
Pressing this button initiates high-output, dual chamber asynchronous pacing. Can be activated anytime:if pacer is locked/unlocked, on/off.
Down
The color of the ventricular cable.
Pacemaker fires at a set rate no matter what the intrinsic rate is. Does not sense.
This color indicator lights up each time the pacemaker delivers a paced event.
These are on the patient's left
Function we DO NOT use on PCU/IMCU. Delivers bursts of atrial pacing up to 800 beats per minute. DO NOT USE!
Always have a spare set of these attached to the back of the pacemaker.
This color indicator lights up for instrinsic sensed events,
Pressing and holding this key suspends pacing and sensing up to 10 seconds. Use this to check intrinsic rhythm.
The color of the atrial cable.
This automatically happens 60 seconds after the last adjustment is made or when lock button is pressed.
Use attachment ring on the back of the pacemaker and attach to this, behind the patient.
Must be worn when touching pacer wires.